Home > Moisture Testing > Trident™ >

  Trident™
Hand Held Microwave Moisture Meter
For Fine and Coarse Aggregates

Trident™ Hand Held Microwave Moisture Meter For Fine and Coarse Aggregates
Trident


 
Questions?
Call us for more information!
We also offer an "Add extended warranty" option to your purchased instrument! Just add it to your cart.


Item# Item Name Price Qty Add
T-T-90 Trident Complete System
Includes Meter, Sensor Probe, Sensor Cable and Software
$1,957.00
T-T-90-ext Trident Moisture Meter Extended Warranty
$453.00
T-T-92 Trident Sensor
$1,299.00
T-T-93 Trident Meter
$1,642.00
Check the items you wish to purchase, then click


Specifications::


James Trident

A microwave meter for rapid determination of moisture content in sand and other fine and coarse aggregrates.


Trident microwave moisture meter for sand and aggreagates



Purchasing Notes

  • The T-T-90 Trident™ is for use only with sand and aggregates, for soils please review the Aquaprobe


Features & Benefits

  • Fast and easy to use; simply insert the prongs into the sand or aggregate being tested

  • Accurate

  • Completely Portable

  • Easy to Read Display


Product Information

A Microwave meter for rapid determination of moisture content in sand and other fine and coarse aggregates.

The Trident™ T-T-90 utilizes the latest microwave and microprocessor technology to measure moisture content in various fine and coarse-grained materials. The prongs of the probe are inserted into the material to be tested and the percentage of moisture content is instantaneously shown on the easy to read display.

An average of five to ten readings is normally taken in order to ensure a valid result. This output is converted by the integrated microprocessor and the moisture content is displayed directly as a percentage of dry weight. The Trident™ comes calibrated for both sand and aggregate, and can be programmed by the user with up to ten different materials.

Trident PC Software

The Trident™ can store more than 150 readings - complete with time and date for future reference. Data can be recalled via USB interface to a personal computer.


Specifications

Specifications
Moisture Range: 0 - 20% by Dry Weight
Frequency: 50 Mhz
Power:

4 AA Batteries

Display: 4 x 16 Char. Trans - reflective
Data Link: USB
Weight: 4 lbs. (1.8 kg)

Sales Numbers
T-T-90: Trident™ Microwave Moisture Meter
T-T-92: Trident™ Microwave Sensor
T-T-93: Trident™ Microwave Meter

Downloadable Documents


Calibration Procedures

  • Dry sample in low heat using oven. It is suggested to weight sample prior to drying.
  • When sample has lost all moisture, calibration can be started.
    Weight dried sample
  • Add water by 2 percent increments.
    i.e. 4,000 grams – 2 percent water = 80 grams of water
  • Turn T-90 on
  • Using up or down arrows, scroll through the menu selection until Direct mode appears on the T-90 display. (It is recommended to erase all prior data saved in the T-90. (T-90 manual page 11))
  • Take 10 readings for each percentage. Store each reading by pressing enter.
  • After T-90 has hit full saturation (direct reading 409.5 percent), uploading data to C. Software is necessary.
  • Uploading can be done by connecting the Trident™ to the P.C. using the supplied RS232 cable.
  • Use instructions in the T-90 manual pages 14-17 to successfully accomplish calibration.
  • Once analyzation has been completed with the T-90 P.C. Software, the Offset and the Gain will be displayed.
    Write down these two values.
  • Return to the T-90 and enter a user program. (T-90 manual page 7 “Change user program”)
    Once the gain and offset have been enterd in the T-90, calibration has been completed.